Remote Control Commands
R&S
®
SMB100A
413
Operating Manual 1407.0806.32 ─ 21
Manual operation:
See
"RDS Program Service Name - Stereo Modulation"
[:SOURce]:STEReo:EXTernal:IMPedance
<Impedance>
Selects the input impedance for the external analog audio signal inputs L and R.
Parameters:
<Impedance>
600 | 600Ohm | 100000 | 100kOhm | 100000Ohm
*RST:
100000
Example:
SOUR:STER:EXT:IMP 600Ohm
selects 600 OHM as the impedance for the external analog
audio signals.
Options:
R&S
SMB-B5
Manual operation:
See
"External R/L Impedance - Stereo Modulation"
[:SOURce]:STEReo:MMF
<Mmf>
Sets the maximum possibly used modulation frequency. This setting is only effective
for external modulation source and activated preemphasis. It prevents over modulation
but result in a decreased s/n ratio.
Parameters:
<Mmf>
integer
Range:
1000 to 18000
*RST:
1000
Example:
SOUR:STER:MMF 2000
sets a maximum modulation frequency of 2 kHz.
Options:
R&S
SMB-B5
Manual operation:
See
"Max Modulation Freq- Stereo Modulation"
[:SOURce]:STEReo:PILot:PHASe
<Phase>
Sets the phase of the pilot tone in degrees, in relation to the 38 kHz carrier signal of
the receiver. For a correct demodulation, the pilot tone must be in phase with the 38
kHz carrier.
Parameters:
<Phase>
float
Range:
-5 to 5
Increment: 0.1
*RST:
0
Example:
SOUR:STER:PIL:PHAS .2DEG
decreases pilot tone quality by adding a phase difference of 0.2
degrees between pilot signal and receiver carrier signal.
SOURce Subsystem