Remote Control Commands
R&S
®
SMB100A
338
Operating Manual 1407.0806.32 ─ 21
[:SOURce<hw>]:FM:EXTernal:DEViation
<Deviation>
Sets the deviation of the external frequency modulation signal in Hz. The maximum
deviation depends on the set RF frequency and the selected modulation mode (see
data sheet).
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command
Parameters:
<Deviation>
float
Range:
see data sheet
Increment: 0.01
*RST:
1000
Example:
FM:EXT:DEV 3kHz
sets 3 kHz deviation to the frequency modulation signal.
Manual operation:
See
[:SOURce<hw>]:FM:INTernal:DEViation
<Deviation>
Sets the deviation of the internal frequency modulation signal in Hz.
The sum of the deviations of all active frequency modulation signals may not exceed
the total value set with command
Parameters:
<Deviation>
float
Range:
0 to dynamic
Increment: 0.01
*RST:
1E3
Example:
FM:INT1:DEV 2E3
sets 2 kHz deviation for the freqeuncy modulation signal.
Manual operation:
See
[:SOURce<hw>]:FM:MODE
<Mode>
Selects the mode for the frequency modulation.
SOURce Subsystem