Remote Control Commands
R&S
®
SMA100B
364
User Manual 1178.3834.02 ─ 03
Example:
SENS1:TYPE?
Queries the type of sensor.
Response: NRP-Z21
The R&S NRP-Z21 sensor is used.
Usage:
Query only
Manual operation:
See
"Sensor type and serial number"
:SENSe<ch>[:POWer]:ZERO
Performs zeroing of the sensor.
Zeroing is required after warm-up, i.e. after connecting the sensor.
Note:
Switch off or disconnect the RF power source from the sensor before zeroing.
We recommend that you zero in regular intervals (at least once a day), if:
●
The temperature has varied more than about 5 °C.
●
The sensor has been replaced.
●
You want to measure very low power.
Example:
SENS1:ZERO
Executes zeroing.
Usage:
Event
Manual operation:
See
12.15 SOURce Subsystem
The
SOURce
subsystem contains the commands for configuring the digital and analog
signals.
Common suffixes
The following common suffixes are used in remote commands:
Suffix
Value range
Description
SOURce<hw>
[1]
●
SOURce[1] = RF output
(optional keyword)
●
LF output =
SOURce:LFOutput
(
SOURce
is optional keyword)
............................................................................365
........................................................................388
.........................................................................395
................................................................................... 402
............................................................................. 403
.....................................................................................416
................................................................................. 429
SOURce Subsystem