Remote Control Commands
R&S
®
SMA100B
309
User Manual 1178.3834.02 ─ 03
Determines whether the contents of the
ENABle
registers are preserved or reset when
the instrument is switched on. Thus a service request can be triggered when the instru-
ment is switched on, if the status registers ESE and SRE are suitably configured. The
query reads out the contents of the "power-on-status-clear" flag.
Parameters:
<Action>
0 | 1
0
The contents of the status registers are preserved.
1
Resets the status registers.
*RCL
<Number>
Recall
Loads the instrument settings from an intermediate memory identified by the specified
number. The instrument settings can be stored to this memory using the command
with the associated number.
It also activates the instrument settings which are stored in a file and loaded using the
MMEMory:LOAD <number>, <file_name.extension>
command.
Manual operation:
See
*RST
Reset
Sets the instrument to a defined default status. The default settings are indicated in the
description of commands.
The command is equivalent to
SYSTem:PRESet
.
Usage:
Setting only
Manual operation:
See
*SAV
<Number>
Save
Stores the current instrument settings under the specified number in an intermediate
memory. The settings can be recalled using the command
with the associated
number.
To transfer the stored instrument settings in a file, use the command
.
Manual operation:
See
Common Commands