Remote Control Commands
R&S
®
SGT100A
474
User Manual 1176.8674.02 ─ 07
:STATus:PRESet
<Preset>
Resets the status registers. All PTRansition parts are set to FFFFh (32767), i.e. all
transitions from 0 to 1 are detected. All NTRansition parts are set to 0, i.e. a transition
from 1 to 0 in a CONDition bit is not detected. The ENABle parts of STATus:OPERa-
tion and STATus:QUEStionable are set to 0, i.e. all events in these registers are not
passed on.
Parameters:
<Preset>
string
Example:
STAT:PRES
resets the status registers.
:STATus:QUEStionable:CONDition
<Condition>
Queries the content of the CONDition part of the STATus:QUEStionable register. This
part contains information on the action currently being performed in the instrument.
The content is not deleted after being read out since it indicates the current hardware
status.
Parameters:
<Condition>
string
Example:
:STATus:QUEStionable:CONDition?
queries the Status:Questionable:Condition register.
:STATus:QUEStionable:ENABle
<Enable>
Sets the bits of the ENABle part of the STATus:QUEStionable register. This setting
determines which events of the Status-Event part are enabled for the sum bit in the
status byte. These events can be used for a service request.
Parameters:
<Enable>
string
Example:
STAT:OPER:ENAB 1
problems when performing an adjustment cause an entry to be
made in the sum bit.
:STATus:QUEStionable[:EVENt]
<Event>
Queries the content of the EVENt part of the
STATus:QUEStionable
register. This
part contains information on the actions performed in the instrument since the last
readout. The content of the EVENt part is deleted after being read out.
Parameters:
<Event>
string
Example:
STAT:QUES:EVEN?
queries the Status:Questionable:Event register.
STATus Subsystem