![R&S RTO6 User Manual Download Page 2538](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646462538.webp)
Remote control commands
R&S
®
RTO6
2538
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
ACQuire:ARESet:TIME
<EnvelopeTimeout>
Defines the time after which the envelope and average evaluation restarts.
The setting is relevant if
TIME
.
Parameters:
<EnvelopeTimeout> Range:
0.1 to 10000
Increment: 0.01
*RST:
0.1
Default unit: s
Usage:
Asynchronous command
ACQuire:ARESet:COUNt
<NofWaveforms>
Defines the number of acquired waveforms after which the envelope and average eval-
uation restarts.
The setting is relevant if
WFMS
.
Parameters:
<NofWaveforms>
Range:
2 to 16777215
Increment: 10
*RST:
1000
Usage:
Asynchronous command
POWer:SLEWrate:GATE:STARt
<T0>
POWer:SLEWrate:GATE:STOP
<T1>
Sets the value for the cursor.
Parameters:
<T1>
POWer:SLEWrate:REPort:ADD
Adds the result to the report list.
Usage:
Event
POWer:SLEWrate:RESult:ACTual?
<MeasType>
POWer:SLEWrate:RESult:AVG?
<MeasType>
POWer:SLEWrate:RESult:EVTCount?
<MeasType>
POWer:SLEWrate:RESult:NPEak?
<MeasType>
POWer:SLEWrate:RESult:PPEak?
<MeasType>
POWer:SLEWrate:RESult:RMS?
<MeasType>
Power analysis (option R&S
RTO6-K31)