![R&S RTO6 User Manual Download Page 1335](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461335.webp)
Remote control commands
R&S
®
RTO6
1335
User Manual 1801.6687.02 ─ 05
Parameters:
<MaxAcqCnt>
Range:
1 to 16777215
Increment: 10
*RST:
1
Usage:
Asynchronous command
ACQuire:ARESet:IMMediate
Forces the immediate restart of the envelope and average calculation for all wave-
forms.
Usage:
Event
Asynchronous command
ACQuire:ARESet:MODE
<ArtmRst>
Defines when the envelope and average evaluation restarts.
Parameters:
<ArtmRst>
NONE | TIME | WFMS
TIME
Restarts the envelope and average calculation after the time
defined with
.
WFMS
Restarts the envelope and average calculation after a number of
acquired waveforms defined with
on page 1335.
*RST:
NONE
Usage:
Asynchronous command
ACQuire:ARESet:TIME
<EnvelopeTimeout>
Defines the time after which the envelope and average evaluation restarts.
The setting is relevant if
TIME
.
Parameters:
<EnvelopeTimeout> Range:
0.1 to 10000
Increment: 0.01
*RST:
0.1
Default unit: s
Usage:
Asynchronous command
ACQuire:ARESet:COUNt
<NofWaveforms>
Defines the number of acquired waveforms after which the envelope and average eval-
uation restarts.
Acquisition and setup