Remote control commands
R&S
®
RTO6
2528
User Manual 1801.6687.02 ─ 05
Parameters:
<Frequency>
F50 | F60 | F360 | F400 | F650 | F800 | NFF650 | WFF800 |
FCUS
NFF650: 360 to 650 Hz
WFF800: 360 to 800 Hz
FCUS: user-defined frequency to be set using
*RST:
F50
POWer:QUALity:FCUS
<CustomFrequency>
Sets the user-defined frequency if
FCUS
.
Parameters:
<CustomFrequency> Range:
1 to 5000
Increment: 1
*RST:
16.666
Default unit: Hz
Firmware/software:
Version 2.70
POWer:QUALity:REPort:ADD
Adds the result to the report list.
Usage:
Event
POWer:QUALity:RESult:CURRent:CREStfactor?
POWer:QUALity:RESult:CURRent:FREQuency?
POWer:QUALity:RESult:CURRent:PEAK?
POWer:QUALity:RESult:CURRent:RMS?
POWer:QUALity:RESult:POWer:APParent?
POWer:QUALity:RESult:POWer:PFACtor?
POWer:QUALity:RESult:POWer:PHASe?
POWer:QUALity:RESult:POWer:REACtive?
POWer:QUALity:RESult:POWer:REALpower?
POWer:QUALity:RESult:VOLTage:CREStfactor?
POWer:QUALity:RESult:VOLTage:FREQuency?
POWer:QUALity:RESult:VOLTage:PEAK?
POWer:QUALity:RESult:VOLTage:RMS?
Returns the value of the respective result.
Return values:
<Value>
Range:
Depends on the measured quantity.
Default unit: Depends on the measured quantity.
Usage:
Query only
Power analysis (option R&S
RTO6-K31)