Remote control commands
R&S
®
RTO6
2474
User Manual 1801.6687.02 ─ 05
Parameters:
<IQInputMode>
BASeband | RFIF
Baseband or intermediate frequency (RFIF)
*RST:
BASeband
Usage:
Asynchronous command
IQ:RBWidth
<RelBw>
Sets the bandwidth factor to define the filter bandwidth.
Filter BW = Relative BW * Sample rate
Parameters:
<RelBw>
Range:
0.04 to 0.8
Increment: 0.01
*RST:
0.8
Usage:
Asynchronous command
IQ:BWIDth?
Returns the filter bandwidth.
See also:
Return values:
<AbsBw>
Range:
40 to 8E+9
*RST:
4E+9
Default unit: Hz
Usage:
Query only
Asynchronous command
IQ:SRATe
<SampleRate>
Sets the required sample rate of the output I/Q data.
Parameters:
<SampleRate>
Range:
1 to 10E+9
Increment: 1
*RST:
10E+9
Default unit: Sa/s
Usage:
Asynchronous command
IQ:RLENgth
<RecordLength>
Sets the required record length of the output I/Q data. The resulting acquisition time of
the I/Q data is:
Acquisition time = Record length / Sample rate
I/Q software interface (option R&S
RTO6-K11)