Remote control commands
R&S
®
RTO6
2023
User Manual 1801.6687.02 ─ 05
TRIGger<m>:ETHernet:FRAMe:DMIN
<DestAddrPattMin>
Specifies a destination address pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<DestAddrPattMin>
TRIGger<m>:ETHernet:FRAMe:SCONdition
<SrcAddrOperator>
Sets source address operator to trigger on a specific pattern or a range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrOperator>
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less than or equal, Greater Than,
Greater than or equal. These conditions require a destination
address to be set with
TRIGger<m>:ETHernet:FRAMe:SMIN
.
INRange | OORange
In range, Out of range. Set the minimum and maximum value of
the range with
TRIGger<m>:ETHernet:FRAMe:SMIN
and
TRIGger<m>:ETHernet:FRAMe:SMAX
*RST:
EQUal
TRIGger<m>:ETHernet:FRAMe:SMAX
<SrcAddrPattMax>
Sets the end value of a range, if the data condition is set to
INRange
or
OORange
.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMax>
TRIGger<m>:ETHernet:FRAMe:SMIN
<SrcAddrPattMin>
Specifies a source address pattern, or sets the start value of a pattern range.
Suffix:
<m>
.
1..3
Parameters:
<SrcAddrPattMin>
Protocols