Remote control commands
R&S
®
RTO6
1759
User Manual 1801.6687.02 ─ 05
BUS<m>:CMSB:NRZ:ENAPolarity
<PolarityEnable>
Sets whether the transmitted enable signal is active when the voltage is below the
threshold (
ENALow
) or higher than it (
ENAHIgh
).
Suffix:
<m>
.
1..4
Parameters:
<PolarityEnable>
ENALow | ENAHigh
*RST:
ENAHigh
BUS<m>:CMSB:NRZ:POLarity
<PolarityData>
Selects the polarity of the custom serial bus data signal in NRZ coding standards.
Suffix:
<m>
.
1..4
Parameters:
<PolarityData>
ACTLow | ACTHigh
ACTLow
Active low: the value "1" is represented by a voltage below the
threshold
ACTHigh
Active high: the value "1" is represented by a voltage above the
threshold
*RST:
ACTHigh
BUS<m>:CMSB:NRZ:THReshold:CLCK
<ThresholdClock>
Sets the threshold for the clock signal digitization in the NRZ Clocked coding standard.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdClock>
Range:
-25 to 25
Increment: 0.1
*RST:
2
Default unit: V
BUS<m>:CMSB:NRZ:HYSTeresis:CLCK
<HystClock>
Sets the hysteresis size on the clock channel.
Suffix:
<m>
.
1..4
Protocols