![R&S RTO6 User Manual Download Page 1618](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461618.webp)
Remote control commands
R&S
®
RTO6
1618
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:WIDTh:RANGe
<SearchName>,<RangeMode>
SEARch:TRIGger:WIDTh:RANGe?
<SearchName>
Selects how the range of a pulse width is defined in relation to the width and delta (see
Parameters:
<RangeMode>
WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers on pulses inside a given range. The range of the pulse
width is defined by "Width" and "
±
Delta".
OUTSide
Triggers on pulses outside a given range. The range definition is
the same as for "Within" range.
SHORter
Triggers on pulses shorter than the given "Width".
LONGer
Triggers on pulses longer than the given "Width".
*RST:
WITHin
Parameters for setting and query:
<SearchName>
Search definition
Example:
See
Chapter 24.3.4.1, "Searching for a pulse of specified width"
Usage:
Asynchronous command
SEARch:TRIGger:WIDTh:WIDTh
<SearchName>,<Width>
SEARch:TRIGger:WIDTh:WIDTh?
<SearchName>
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "
±
Delta".
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
.
Parameters:
<Width>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
5E-9
Default unit: s
Parameters for setting and query:
<SearchName>
Search definition
Example:
See
Chapter 24.3.4.1, "Searching for a pulse of specified width"
Usage:
Asynchronous command
Search