Remote control commands
R&S
®
RTO6
1611
User Manual 1801.6687.02 ─ 05
WITHin
Triggers on pulse intervals inside a given range. The range is
defined by "Interv. width" and "
±
Delta".
OUTSide
Triggers on intervals outside a given range. The range definition
is the same as for "Within" range.
SHORter
Triggers on intervals shorter than the given "Interv. width".
LONGer
Triggers on intervals longer than the given "Interv. width".
*RST:
OUTSide
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:INTerval:WIDTh
<SearchName>,<Width>
SEARch:TRIGger:INTerval:WIDTh?
<SearchName>
Defines the time between two pulses.
Parameters:
<Width>
Range:
100E-12 to 864
Increment: 100E-9
*RST:
5E-9
Default unit: s
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
24.15.6
Runt search conditions
..................................................................................... 1611
..................................................................................1612
................................................................................... 1612
.................................................................................... 1613
SEARch:TRIGger:LEVel:RUNT:LOWer
...........................................................................1613
SEARch:TRIGger:LEVel:RUNT:UPPer
........................................................................... 1613
SEARch:TRIGger:RUNT:DELTa
<SearchName>,<WidthDelta>
SEARch:TRIGger:RUNT:DELTa?
<SearchName>
Defines a range around the given runt width.
Search