Remote control commands
R&S
®
RTO6
1609
User Manual 1801.6687.02 ─ 05
SEARch:TRIGger:GLITch:POLarity
................................................................................ 1609
................................................................................. 1609
...................................................................................1609
SEARch:TRIGger:GLITch:POLarity
<SearchName>,<Polarity>
SEARch:TRIGger:GLITch:POLarity?
<SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.
Parameters:
<Polarity>
POSitive | NEGative | EITHer
See
Chapter 24.4.4, "Polarity parameter"
*RST:
POSitive
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:GLITch:RANGe
<SearchName>,<RangeMode>
SEARch:TRIGger:GLITch:RANGe?
<SearchName>
Selects which glitches are identified: shorter or longer than the specified width (see
Parameters:
<RangeMode>
SHORter | LONGer
*RST:
SHORter
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
SEARch:TRIGger:GLITch:WIDTh
<SearchName>,<Width>
SEARch:TRIGger:GLITch:WIDTh?
<SearchName>
Sets the length of a glitch. The instrument triggers on pulses shorter or longer than this
value (see also
You need to know the expected pulse widths of the circuit to set the glitch width cor-
rectly.
Parameters:
<Width>
Range:
100E-12 to 1E-3
Increment: 100E-6
*RST:
1E-9
Default unit: s
Parameters for setting and query:
<SearchName>
Search definition
Usage:
Asynchronous command
Search