CALCulate:MATH<m>:FFT:PHASe:SCALe
<PhaseScale>
Defines the scaling unit for phase display.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<PhaseScale>
DEGRees | RADians
*RST:
DEGRees
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:SUPPression
<Suppression>
Enables noise suppression. Phase calculation is restricted to frequencies with a mini-
mum magnitude, the threshold value (see
Suffix:
<m>
.
1..8
math waveform
Parameters:
<Suppression>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:THReshold
<SupprThres>
Defines the minimum frequency magnitude for which phases are calculated. This set-
ting is only available if
CALCulate:MATH<m>:FFT:PHASe:SUPPression
is set to
"ON".
Suffix:
<m>
.
1..8
math waveform
Parameters:
<SupprThres>
Range:
-180 to 180
Increment: 0.1
*RST:
0
Default unit: dBm
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:PHASe:UNWRap
<Unwrap>
If enabled, phase shifts due to a limitation of the value range are eliminated.
Spectrum analysis