Remote control commands
R&S
®
RTO6
1559
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see
Parameters:
<Value>
Example:
REFLevel2:ABSolute:ULEVel 0.01
Sets the upper reference level for C1W1 to 10 mV. C1W1 corre-
sponds to suffix number 2.
Example:
See:
"Manual reference level definition using absolute values"
Usage:
Asynchronous command
REFLevel<m>:ABSolute:LLEVel
<Value>
The lower reference level, required e.g. to determine a fall.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see
Parameters:
<Value>
Example:
REFLevel2:ABSolute:LLEVel 0.001
Sets the lower reference level for C1W1 to 1 mV. C1W1 corre-
sponds to suffix number 2.
Example:
See:
"Manual reference level definition using absolute values"
Usage:
Asynchronous command
REFLevel<m>:RELative:UPPer
<Value>
Sets the upper relative reference level if
USER
.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 118, 151 to 157,
173 to 179
Source waveform of the measurement, see
Parameters:
<Value>
Percentage of the high signal level.
Automatic measurements