Remote control commands
R&S
®
RTO6
1548
User Manual 1801.6687.02 ─ 05
MEASurement<m>:GATE:CURSor
<Cursorset>
Selects the cursor set to be used for measurement gating. The gate area is defined by
the cursor lines.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<Cursorset>
CURSOR1 | CURSor1 | CURSOR2 | CURSor2 | CURSOR3 |
CURSor3 | CURSOR4 | CURSor4
CURSOR1 = CURSor1, CURSOR2 = CURSor2, CURSOR3 =
CURSor3, CURSOR4 = CURSor4
*RST:
CURSOR1
Usage:
Asynchronous command
MEASurement<m>:GATE:CCOupling
<CursorCoupling>
Enables the cursor coupling for automatic measurements.
Select the cursor set to be used with
.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<CursorCoupling>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
MEASurement<m>:GATE:ZCOupling
<ZoomCoupling>
If enabled, the gate area is defined identically to the zoom area for the zoom diagram.
If enabled, define the zoom area to be used as gate with
.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<ZoomCoupling>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Automatic measurements