![R&S RTO6 User Manual Download Page 1534](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461534.webp)
Remote control commands
R&S
®
RTO6
1534
User Manual 1801.6687.02 ─ 05
Usage:
Event
Asynchronous command
MEASurement<m>:MULTiple
<MultiMeas>
The measurement is performed repeatedly if the measured parameter occurs several
times inside the acquisition or defined gate.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<MultiMeas>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
MEASurement<m>:MNOMeas
<MaxMeasPerAcq>
Sets the maximum number of measurements per acquisition if repeated measurement
is enabled (
ON
).
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<MaxMeasPerAcq>
Range:
2 to 1000000
Increment: 1
*RST:
1000
Usage:
Asynchronous command
MEASurement<m>:STATistics:HISTogram
<ShowHistogram>
Displays a histogram of the statistical results. Enabling the histogram enables also the
calculation and display of statistics for the measurement results if statistics were dis-
abled. the histogram shows the cumulative occurence distribution of mean measure-
ment results in a graphic.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<ShowHistogram>
ON | OFF
*RST:
OFF
Automatic measurements