Remote control commands
R&S
®
RTO6
1493
User Manual 1801.6687.02 ─ 05
MEASurement<m>:DETThreshold
<SignDetectThres>
Defines the value above which measurement results are displayed. Values beneath the
threshold are considered to be noise and they are ignored.
Suffix:
<m>
.
1..10
irrelevant
Parameters:
<SignDetectThres>
Range:
0 to 50
Increment: 1
*RST:
5
Default unit: %
Usage:
Asynchronous command
MEASurement<m>:AMPTime:ALEVel
<AreaLevel>
Defines the reference level used to integrate the waveform.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<AreaLevel>
Range:
-100E+24 to 100E+24
Increment: 0
*RST:
0
Default unit: V
Usage:
Asynchronous command
MEASurement<m>:AMPTime:PFSLope
<PeriodSlope>
Selects the slope direction for frequency and period measurements.
Suffix:
<m>
.
1..10
See
"Selection of the measurement group: MEASurement<m>"
Parameters:
<PeriodSlope>
FIRSt | POSitive | NEGative | EITHer
POSitive | NEGative
Measures the time between rising or falling edges, respectively.
EITHer
In multiple measurements, the time is measured both between
rising edges and between falling edges.
In single measurements. the first edge is taken for the measure-
ment.
Automatic measurements