Remote control commands
R&S
®
RTO6
1391
User Manual 1801.6687.02 ─ 05
SETHold
SETHold: analyzes the relative timimg between a data signal
and the synchronous clock signal. For trigger settings, see
Chapter 24.9.10, "Setup & Hold"
ANEDge
Edge trigger for external trigger input. Only available if the trig-
ger source is the external trigger input. This trigger type uses the
analog input signal. For trigger settings, see
SERPattern
Serial pattern for signals with serial data patterns in relation to a
clock signal. For trigger settings, see
NFC
Specific trigger for Near Field Communication testing, requires
option R&S
RTO6--K11. For NFC trigger settings, see
CDR
Trigger on the edges of a recovered clock signal, requires option
R&S
RTO6--K13. The CDR also uses the trigger level as thresh-
old to detect signal edges. For CDR trigger settings, see
ter 24.22.4, "Clock data recovery (hardware-based, option
R&S
*RST:
EDGE
Usage:
Asynchronous command
TRIGger<m>:LEVel<n>[:VALue]
<Level>
Sets the trigger level for the specified event and source.
If the trigger source is serial bus, the trigger level is set by the thresholds in the proto-
col configuration.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n>
1..31
Indicates the trigger source:
1...4 = channel 1 to 4
5 = external trigger input
6...11 = not available
12...27 = R&S
RT-ZVC input channels
28...31 = differential signals Diff1, Diff2 and Common1, Com-
mon2
Trigger