Remote control commands
R&S
®
RTO6
1369
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:OVERload:RSTO
<Settings>
Resets the overload indication at the probe.
In internal shunt mode (
INTShunt
), chooses
whether to keep the operation range adjusted by the R&S
RT-ZVC during overload or
to restore the original settings for the given current channel.
In external shunt mode (
EXTShunt
), there is
no adjustment of the operation range during overload.
Suffix:
<m>
.
1..2
<n>
1..4
Setting parameters:
<Settings>
ADJusted | ORIGinal
ADJusted
The operation range that is automatically adjusted by the probe
during overload is kept.
In internal shunt mode, during an overload at a current channel,
the probe switches automatically to the next higher range. If the
10A range is also exceeded, the amperemeter switches to exter-
nal shunt mode to protect the probe against permanent damage.
ORIGinal
The original operation range is restored as before the overload
and the adjustment of the operation range.
Usage:
Setting only
Asynchronous command
ZVC:Z<m>:I<n>:OVERload:VALue?
Queries if an overload of the current channel was detected.
Suffix:
<m>
.
1..2
<n>
1..4
Return values:
<Overload>
ON | OFF
*RST:
OFF
Usage:
Query only
Asynchronous command
ZVC:Z<m>:I<n>:POSition
<VertPosi>
Sets the vertical position of the indicated current channel as a graphical value.
Acquisition and setup