Remote control commands
R&S
®
RTO6
1331
User Manual 1801.6687.02 ─ 05
Parameters:
<RecordLength>
Number of recorded waveform points.
Range:
1000 to 1000000000
Increment: 2
*RST:
1000
Default unit: pts
Usage:
Asynchronous command
TIMebase:ROLL:ENABle
<Mode>
Activates the automatic roll mode.
Parameters:
<Mode>
AUTO | OFF
AUTO: the instrument activates the roll mode under specific con-
ditions.
See also
*RST:
AUTO
Usage:
Asynchronous command
TIMebase:ROLL:STATe?
Returns the status of the roll mode.
Return values:
<State>
ON | OFF
*RST:
OFF
Usage:
Query only
Asynchronous command
TIMebase:ROLL:MTIMe
<MinHorizGain>
The roll mode is enabled automatically if the acquisition time exceeds the given value,
and if
Parameters:
<MinHorizGain>
Treshold value for roll mode enabling.
Range:
1 to 600
Increment: 1
*RST:
10
Default unit: s
Usage:
Asynchronous command
Acquisition and setup