![R&S RTO6 User Manual Download Page 1317](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461317.webp)
Remote control commands
R&S
®
RTO6
1317
User Manual 1801.6687.02 ─ 05
MTESt:LABel
<LabelState>
Switches the display of the mask test name on or off.
To change the name of the mask test, use
Parameters:
<LabelState>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
MTESt:REName
<MaskTestName>, <NewName>
MTESt:REName?
<MaskTestName>
Changes the name of the mask test.
Parameters:
<NewName>
String with the new mask test name.
Parameters for setting and query:
<MaskTestName>
String with the existing mask test name.
MTESt:HIGHlight:INFinite
<State>
If ON, the mask hits are highlighted for an unlimited period of time.
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
MTESt:HIGHlight:STATe
<HighlightHits>
If ON, the mask hits are highlighted on the screen. You can define the color and the
time of the hit display.
Parameters:
<HighlightHits>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
MTESt:HIGHlight:TIME
<HighlightTime>
Sets the time how long the mask hits are highlighted.
Instrument setup