![R&S RTO6 User Manual Download Page 1242](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461242.webp)
Network operation and remote control
R&S
®
RTO6
1242
User Manual 1801.6687.02 ─ 05
●
hislip0
indicates the HiSLIP protocol
●
inst0
is the default LAN device name. VISA supports several devices running on
the instrument. On R&S
RTO6, only one device is configured, so the LAN device
name can be omitted.
●
INSTR
specifies a VISA resource of the type INSTR. By default, the VISA resource
name control is set to the INSTR class.
Example: HiSLIP
IP address is
192.1.2.3
: the valid resource string is:
TCPIP::192.1.2.3::hislip0
Instrument name is
RSRT1
: the valid resource string is:
TCPIP::RSRT1::hislip0
.
DNS host name is
RTO6-123456
: the valid resource string is:
TCPIP::RTO6-123456::hislip0
.
Example: VXI-11
IP address is
192.1.2.3
: the valid resource string is:
TCPIP::192.1.2.3
Instrument name is
RSRT1
: the valid resource string is:
TCPIP::RSRT1
.
DNS host name is
RTO6-123456
: the valid resource string is:
TCPIP::RTO6-123456
.
See also:
●
Find IP address: "Menu" > "Settings" > "System" tab, see
●
Chapter 22.2.2, "Assigning the IP address"
22.5.1.2
GPIB interface (IEC/IEEE bus interface)
A GPIB interface is provided with option R&S
RTO6-B10.
To control the instrument remotely via GPIB, the instrument and the controlle must be
connected by a GPIB bus cable. The total cable length is restricted to a maximum of
15 m; the cable length between two instruments should not exceed 2m.
GPIB instrument address
The GPIB address is set on the instrument, see
Chapter 22.5.2.1, "GPIB settings"
22.5.2
Remote settings
The settings on this tab are required for remote control of the instrument via a connec-
ted computer.
22.5.2.1
GPIB settings
Access: "Menu" > "Settings" > "System" > "Remote"
Displays the GPIB settings of the instrument.
It requires an optional GPIB bus card, option R&S
RTO6-B10.