Deembedding (option R&S
RTO6-K121)
R&S
®
RTO6
1201
User Manual 1801.6687.02 ─ 05
Enable
Activates the deembedding - the correction of parasitic effects of the measurement
setup on the measured signal.
Remote command:
Effective bandwidth
Sets the maximum bandwidth until which the signal is corrected. This maximum value
is the minimum bandwidth value of the probe bandwidth, tip module bandwidth
(R&S
RT
‑
ZM) and oscilloscope bandwidth, and can not be higher than the highest fre-
quency in a used S-parameter file. The setting is also coupled with the digital filter
bandwidth if the cut-off frequency of the digital filter is at least 1
GHz.
Remote command:
Delay compensation
Enables the compensation for the group delay of the complete measurement setup. If
enabled, the calibration time reference point is the start point of the measurement
setup. If disabled, the calibration time reference point is the oscilloscope's channel
input.
Remote command:
Meas setup definition
Simulates the complete measurement setup with all components.
Deembedding settings