Mixed signal option (MSO, R&S
RTO6-B1)
R&S
®
RTO6
1007
User Manual 1801.6687.02 ─ 05
Width settings
The width settings "Range", "Width" and "
±
Delta" appear if the timing mode is set to
"Width".
Range ← Width settings
Selects how the range of a pulse width is defined:
"Within"
Triggers when the pattern comes false inside a given time range. The
time limit is defined by
Width
±
Delta
.
"Outside"
Triggers when the pattern comes false before or after the given time
range. The time limit definition is the same as for "Within" range.
"Shorter"
Triggers when the pattern comes false before the given "Width" has
expired.
"Longer"
Triggers when the pattern comes false after the given "Width" has
expired.
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh:RANGe
Width ← Width settings
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
time limit, respectively.
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "
±
Delta".
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh[:WIDTh]
±Delta ← Width settings
Defines a range around the given width value.
The combination "Range" = Within and "
±
Delta" = 0 triggers on pulses with a pulse
width that equals "Width".
The combination "Range" = Outside and "
±
Delta" = 0 means to trigger on pulse widths
≠
"Width".
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh:DELTa
14.3.1.8
Serial pattern
The serial pattern trigger identifies a serial bit string trigger on a single digital channel,
or for a logical combination of digital channels. The trigger requires a clocked bus; the
bits are read at the specified clock edge. The trigger event occurs at the last clock
edge of the serial bit string.
This trigger type allows you to trigger on specific address or data transmissions in
serial input and output signals.
Trigger