Mixed signal option (MSO, R&S
RTO6-B1)
R&S
®
RTO6
1004
User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:PARallel:SETHold:HTIMe
14.3.1.6
State
The state trigger detects the logical state of several logically combined digital channels
at a given clock edge. The trigger source is a logical combination of digital channels or
a parallel bus. The trigger occurs at the clock edge at which the state condition is true.
Figure 14-5: State trigger settings for trigger source = parallel bus
Clock source
Selects the digital channel of the clock signal.
Remote command:
TRIGger<m>:PARallel:SETHold:CSOurce[:VALue]
TRIGger<m>:PARallel:STATe:CSOurce:VALue
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue]
Clock edge
Sets the edge of the clock signal. The crossing of the clock edge and the logical
threshold defines the time at which the logical states and the bus value are analyzed.
Remote command:
TRIGger<m>:PARallel:STATe:CSOurce:EDGE
Channel states
Available for "Trigger Source"= "Par. Bus1/2/3/4"
For each digital channel that is used in the bus, set the required state: 1, 0, or X (don't
care).
Trigger