Trying Out the Instrument
R&S
®
RTO2000
55
Getting Started 1332.9719.02 ─ 10
During running display, the "Play" is labeled "Running".
15. Close the quick access "History" dialog box.
The history mode is disabled. The [History] key is no longer illuminated.
6.7
Performing Basic Measurements
Using the R&S
RTO, you can perform and display different measurements simul-
taneously. The color of the results in the result table corresponds with the source
waveform color.
6.7.1
Performing a Cursor Measurement
1. Restore the default signal channel settings: Press the [Preset] and [Autoset]
keys.
2. Tap the "Cursor" icon on the toolbar.
3. Draw a rectangle on the screen where you want to place the cursor lines.
The cursor lines appear in the diagram and the "Cursor Results" box opens.
The measured waveform values at the cursor positions are displayed.
4. You can move the cursor lines in different ways:
● Touch a cursor line and drag it on the screen.
● Tap a cursor line to activate it. Then turn the [Navigation] knob to adjust
the position.
● Enter the position values in the result box.
Performing Basic Measurements