Measuring Procedure
R&S
®
RT-ZC31
30
User Manual 1801.4949.02 ─ 01
5. Remove the power plug of the device from the R&S
RT-ZA13 probe power
supply.
6. Unplug the R&S
RT-ZA13 probe power supply power cord from the electrical
outlet.
Risk of instrument damage
●
To prevent wire breaks, do not pull on the cord to disconnect the output
connector from the waveform measurement instrument. Always grip the
terminator and pull the unlock lever toward you before disconnecting the
connector.
●
When disconnecting the device, grip the power supply plug’s shell. Do
not pull on cords to prevent wiring breaks in the device’s power cord or
damage to R&S
RT-ZA13 probe power supply.
3.5
How to Measure Current Accurately
Retracting and extending the upper jaw can cause an offset voltage of several
millivolts. Perform the steps described below to accurately measure a current.
1. Hold down the DEMAG / AUTO ZERO key on the junction box for about 1
second.
ZERO LED lights up, and demagnetization and automatic zero-adjustment
start. After the completion, the DEMAG / AUTO ZERO LED goes out.
2. Wait for about 5 minutes for the fluctuation of the offset voltage to stabilize.
3. Operate the opening lever of the sensor back and forth to retract and extend
the upper jaw 4 or 5 times.
4. Slide the opening lever toward the lower jaw until the JAW UNLOCKED indi-
cator is hidden.
The upper jaw is locked with the upper and lower sensor heads arranged in
position relative to each other. The JAW UNLOCKED LED goes out.
5. Momentarily press the DEMAG / AUTO ZERO key on the junction box. (Do
not hold down the key).
How to Measure Current Accurately