Remote Control - SCPI
R&S
®
QAR
79
User Manual 1178.7200.02 ─ 04
Example:
//Query unit of the attenuation measurement
MEAS:STAR
MEAS:ATT:UNIT?
Usage:
Query only
MEASurement:BTYPe
<BumperType>
This command sets the bumper type.
Parameters:
<BumperType>
String containing the bumper type name
Example:
//Set the bumper type.
MEAS:BTYP 'RR_L'
MEASurement:NORMalize:REQuired?
This command queries the normalization state of the attenuation measurement.
Return values:
<State>
0
Measurement is normalized.
1
Measurement is not normalized. In that case, you should nor-
malize the measurement before starting it.
Example:
//Query normalization state
MEAS:NORM:REQ?
Usage:
Query only
MEASurement:NORMalize:STARt
This command initiates the normalization of an attenuation measurement.
We recommend to define a timeout of at least 10 seconds to allow for the measure-
ment to finish.
Example:
//Normalize transmission measurement
MEAS:NORM:STAR
Usage:
Event
MEASurement:OK?
This command returns
1
if the measured mean attenuation at a specified location is
below threshold.
Remote Commands