Remote Control Commands
R&S
®
NRX
429
User Manual 1178.5566.02 ─ 07
[SENSe<Sensor>:]INPut:ATTenuation[:VALue]
<value>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>]:INPut:ATTenuation[:
VALue]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<value>
Range:
0.0 to 30.0
*RST:
30.0
Default unit: dB
[SENSe<Sensor>:]INTernal:TRIGger:JITTer:METHod
<method>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
TRIGger<Measurement>[:CHANnel<Channel>]:JITTer:METHod
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<method>
COMPensate | MEASure | NONE
*RST:
COMPensate
[SENSe<Sensor>:][POWer:][AVG:]APERture[:VALue]
<integration_time>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
CALCulate<Measurement>[:CHANnel<Channel>][:POWer][:AVG]:
APERture[:VALue]
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<integration_time>
Range:
8.3e-9 to 30.0
*RST:
0.005
Default unit: s
[SENSe<Sensor>:][POWer:][AVG:]SMOothing:STATe
<state>
Available for R&S NRP2 compatibility. Recommended R&S NRX command:
R&S NRP2 Compatibility
www.allice.de
Allice Messtechnik GmbH