Configuration for All Measurement Types
R&S
®
NRX
48
User Manual 1178.5566.02 ─ 05
Remote command:
UNIT<Measurement>:POWer[:VALue]
Forward Unit
Available for NRT measurements.
Specifies the unit of measurement in forward direction. See
Remote command:
UNIT<Measurement>:POWer[:VALue]
Display Format
Specifies the display format of the measured values.
"Scalar Digital" Available for continuous average, burst average, pulse analysis, time
gate, timeslot, NRT measurements.
Numeric format
"Scalar Ana-
log"
Marker on a scale
"Graphical"
Available for time gate, timeslot measurements.
Measured values are plotted over time.
Remote command:
Auxiliary Values
Available for the graphic displays of continuous average, burst average measure-
ments. Only displayed if the measurement result display is shown in full screen.
Determines which additional information about the measured values is shown in the
display. If you press [1 Trig / Delete], you reset the auxiliary values.
"None"
No additional values are measured.
"Extremes"
Displays the maximum, the minimum and the max-min values since
the search for extreme values has been started. With logarithmic
units, the peak-to-peak value equals the quotient of the measured
values converted into linear units.
"Statistics"
Displays the longterm mean, the standard deviation and the total
number of measurement results that have been evaluated since the
search for statistic values has been started.
Remote command:
Scaling
Specifies the scaling of the display. The available parameters depend on the following
settings:
●
.
●
Display Settings