Remote Control Commands
R&S
®
NRX
331
User Manual 1178.5566.02 ─ 05
INPut<Sensor>:PORT:OFFSet
<offs>
Considers the transmission loss in a cable that connects the desired measurement
point, set by
, and the power sensor.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<offs>
Range:
0.0 to 100.0
*RST:
0.0
Default unit: dB
Manual operation:
See
INPut<Sensor>:PORT:POSition
<pos>
Selects the reference plane. It defines at which sensor port the forward and reverse
power is measured.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<pos>
SOURce | LOAD
SOURce
Source connector of the power sensor
LOAD
Load connector of the power sensor
*RST:
SOURce
Manual operation:
See
See
14.6.8.3
NRT Filter Settings
Further information:
●
Chapter 9.5.3, "NRT Filter Settings"
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt:AUTO[:STATe]
CALCulate<Measurement>[:CHANnel<Channel>]:AVERage:COUNt[:VALue]
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture:MODE
............................332
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:APERture[:VALue]
.......................... 333
CALCulate<Measurement>[:CHANnel<Channel>]:NRT:VBWidth[:VALue]
............................ 333
[SENSe<Sensor>:]BWIDth:VIDeo:FNUMber
.....................................................................333
[SENSe<Sensor>:]BANDwidth:VIDeo:FNUMber
............................................................... 333
Configuring Sensors