Remote Control Commands
R&S
®
NRX
319
User Manual 1178.5566.02 ─ 05
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<mode>
MOVing | REPeat
MOVing
Outputs intermediate values to facilitate early detection of
changes in the measured quantity. In the settled state, that
means when the number of measurements specified by the
average count has been performed, a moving average is output.
REPeat
Specifies that a measurement result is not output until the entire
measurement has been completed. This means that the number
of measurement cycle repetitions is equal to the set average
count. If the average count is large, the measurement time can
be very long.
The average count is set using
CHANnel<Channel>]:TRACe:AVERage:COUNt[:VALue]
*RST:
MOVing
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:AVERage[:STATe]
<value>
Effective for trace measurements.
Enables the filter function of a power sensor. If enabled, the number of measured val-
ues is averaged. This reduces the effect of noise so that more reliable results are
obtained.
The number of measured values is set using
CHANnel<Channel>]:TRACe:AVERage:COUNt[:VALue]
.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:NCORrection[:STATe]
<state>
Enables or disables the noise cancellation.
Configuring Sensors