Remote Control Commands
R&S
®
NRX
306
User Manual 1178.5566.02 ─ 05
CALCulate<Measurement>[:CHANnel<Channel>]:TRACe:ESAMpling
<value>
Effective for trace measurements.
Enables or disables the automatic equivalent sampling that allows for high resolution
measurements.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
*RST:
ON
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>]:SAMPling
<value>
Effective for continuous average measurements.
Sets the sampling rate.
With multipath power sensors, you can set the sampling rate to two different values to
prevent aliasing effects for particular types of modulation signal. Aliasing can occur
because the sampling frequency is located within the video bandwidth, which means
that spectral components near the sampling frequency can cause beating effects. With
changing the sampling rate, the beating effects usually disappear.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
<Channel>
1 to 2
1 = primary sensor, 2 = secondary sensor
Parameters:
<value>
FREQ1 | FREQ2
*RST:
FREQ1
Manual operation:
See
CALCulate<Measurement>[:CHANnel<Channel>][:POWer]:BURSt:DTOLerance
<value>
Effective for burst average measurements.
Defines the end of the burst. If power keeps low for at least this time, the end of the
burst is assumed. Modulation-specific power drops that are shorter than the set value
are ignored.
See also
Configuring Sensors