Remote Control Commands
R&S
®
NRX
302
User Manual 1178.5566.02 ─ 05
..............................................................................................323
.................................................................................326
14.6.1
Setting the Frequency
[SENSe<Sensor>:]FREQuency[:CW]
<frequency>
Sets the carrier frequency of the applied signal. This value is used for frequency-
response correction of the measurement result.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
1.0e9
Default unit: Hz
Manual operation:
See
[SENSe<Sensor>:]FREQuency:FIXed
<frequency>
See
[SENSe<Sensor>:]FREQuency[:CW]
Suffix:
<Sensor>
.
1...128
Parameters:
<frequency>
Range:
0.0 to 110.0e9
*RST:
1.0e9
Default unit: Hz
[SENSe<Sensor>:]FREQuency:TRACk
<state>
Enables or disables the frequency tracker of the power sensor, if available.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
OFF | ON
*RST:
OFF
Configuring Sensors