Remote Control Commands
R&S
®
NRX
268
User Manual 1178.5566.02 ─ 05
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
*RST:
OFF
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
<lower>
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
is disabled.
Sets the lower scale limit for the power indication.
The entry has no unit. The unit corresponds to the selected output unit. If you change
the unit, the entered value remains the same.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<lower>
Range:
-1999.0 to 1999.0
*RST:
0.0
[SENSe<Sensor>:]POWer[:POWer]:RANGe[:UPPer]
<upper>
[SENSe<Sensor>:]POWer:REFLection:RANGe:AUTO
is disabled.
Sets the lower scale limit for the power indication. For further details, see
[SENSe<Sensor>:]POWer[:POWer]:RANGe:LOWer
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<upper>
Range:
-1999.0 to 1999.0
*RST:
1.0
[SENSe<Sensor>:]RRESolution
<rres>
Requires the sensor interface for R&S NRT (R&S NRX-B9).
Sets the measurement resolution, thus effecting the accuracy of the measurement, the
measurement duration, and the number of digits indicated on the display.
Suffix:
<Sensor>
.
101
Parameters:
<rres>
LOW | HIGH
Measurement Settings and Results