Remote Control Commands
R&S
®
NRX
220
User Manual 1178.5566.02 ─ 05
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:INFO?
[<ITEM>]
Only for compatibility reasons.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Query parameters:
<ITEM>
Usage:
Query only
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the number of desired values for the buffered continuous average measurement.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<count>
Range:
1 to 131072
*RST:
1
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average measurement. If enabled, data
blocks rather than single measured values are returned and a higher data rate is ach-
ieved.
Suffix:
<Sensor>
.
1 to 128
Configured sensor connected at: port A = 1, ... , port D = 4, R&S
NRX-B9 = 101, USB and LAN port = 5 to 100
Parameters:
<state>
ON | OFF
*RST:
OFF
[SENSe<Sensor>:]POWer:REFerence
<ref>
Sets the reference value for the relative power indication.
Measurement Settings and Results