Remote Control Commands
R&S
®
NRPxxS(N)
126
User Manual 1177.5079.02 ─ 10
............................................................................................ 126
................................................................................................ 126
.................................................................................. 126
CALibration<Channel>:ZERO:AUTO
................................................................................126
CALibration:DATA
<caldata>
Writes a binary calibration data set in the memory of the sensor.
Parameters:
<caldata>
<block_data>
CALibration:DATA:LENGth?
Queries the length in bytes of the calibration data set currently stored in the flash mem-
ory. Programs that read out the calibration data set can use this information to deter-
mine the capacity of the buffer memory required.
Example:
CAL:DATA:LENG?
Query
57392
Response
Usage:
Query only
CALibration:USER:DATA
<caldata>
Accesses the user calibration data block.
Parameters:
<caldata>
<block_data>
CALibration:USER:DATA:LENGth?
Queries the length of the user calibration data block.
Usage:
Query only
CALibration<Channel>:ZERO:AUTO
<state>
Performs zero calibration.
Turn off all test signals before zeroing. An active test signal during zeroing causes an
error.
While zero calibration is in progress, no queries or other setting commands are
allowed, since the command is synchronous. Any communication attempt can run into
a timeout.
Calibrating/Zeroing the Power Sensor