Instrument Functions
R&S
®
NRP2
160
User Manual 1173.9157.02 ─ 03
4.7.6.1
Internal sensor selftest
► Open the "Sensortest" dialog box in the "System"
→
"Test..." menu.
Fig. 4-92: Sensortest dialog box / page 1
The first page of the "Sensortest" dialog box covers the settings for performing an internal
selftest of the sensor without using the reference signal.
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Selftest internal (with no signal)
Starts a selftest of the sensor. A signal must not be applied.
The instrument indicates the corresponding states to internal self-test directly in the
menu:
●
"...not tested" (prior to testing)
●
"...running" (in progress)
●
"...successful" (outcome)
●
"...failed" (outcome)
Test Report…
Opens the "Sensor Selftest Report" window displaying the results of the internal selftest.
The status bar of the report window shows type and serial number of the tested sensor,
and the overall test result. The results of the individual tests are indicated below.
Fig. 4-93: Sensor Selftest Report dialog box
System settings
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...