Instrument Functions
R&S
®
NRP2
86
User Manual 1173.9157.02 ─ 03
"Single"
"Trace&Statistics" mode only. This setting disables continuous trigger-
ing so that only one trigger event at a time is executed. To enable trig-
gering again, press the (DEL/1TRIG) key. The symbol
appears
in the display (see
chapter 4.2.7, "Trigger sequence control"
"Normal"
Continuous triggering with regular trigger events.
"Auto"
"Trace&Statistics" mode only. This setting automatically starts a mea-
surement if no trigger event has occurred after 300 ms.
SCPI command:
Source
Selects the trigger source for the selected sensor. The measured internal signal ("Inter-
nal") or an external signal is available for triggering. The external trigger signal is either
input at the TRIG IN / OUT2 connector at the rear of the instrument ("External") or it is
supplied by another power sensor which must be configured as a trigger master ("Int.
Master").
Note:
In addition to using an internal or external trigger source, a trigger event can be
forced using the DEL/1TRIG hardkey, see
chapter 4.2.7, "Trigger sequence control"
"Internal"
The trigger event for the selected sensor is derived from the measured
signal.
"External"
The trigger event for the selected sensor is derived from the trigger bus.
The trigger signal on the trigger bus is either input at the TRIG IN /
OUT2 connector at the rear of the instrument or it is provided by another
sensor.
If the trigger signal is input at the TRIG IN / OUT2 connector, the con-
nector must be configured for "Trigger In" in the "System" menu, see
chapter 4.7.2.2, "I/O 2 interface"
If the trigger signal is provided by another sensor, this sensor must be
configured as trigger master, see below, "Int. Master".
"Int. Master"
The selected power sensor is automatically configured as trigger mas-
ter and set to internal triggering. Any previously defined trigger master
is disabled and an external trigger signal at the TRIG IN / OUT2 con-
nector is ignored.
The trigger master is used as the external trigger source for all other
sensors.
SCPI command:
Slope
Sets the polarity of the active slope of the externally or internally applied trigger signal,
see
"
Positive"
The rising edge of the trigger signal is active.
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...