Instrument Functions
R&S
®
NRP2
84
User Manual 1173.9157.02 ─ 03
Offset
Negative dB value as a measure for the shift of the transition region(s).
Modifications are immediately effective after each keystroke (see
SCPI command:
4.2.5 Effect of the RF source
Information related to the reflection coefficient
Γ
of the signal source is provided in the
"
Γ
Correction" dialog box (gamma correction).
Γ
Correction is used to increase measure-
ment accuracy by correcting the effect of source mismatch on the measured value.
4.2.5.1
Sensor
Γ
correction dialog box
► Open the
Γ
correction dialog box in the "Sensor "menu
If "
Γµ
correction" is activated, the measurement accuracy can be increased by specifying
magnitude and phase of source
Γ
.
Fig. 4-19:
Γ
Correction dialog box
A | B | C | D
Tabs for selecting the sensor.
Note:
In remote-control commands, select the sensor with the suffix in command
SENSe<[1]..4>:...
, i.e. choose suffix
1
for channel A, channel
2
for B, and so on.
Source
Γ
Correction
Activates gamma correction in order to achieve higher measurement accuracy.
SCPI command:
Data acquisition and parameters
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...