Remote Control
R&S
®
NGE100B
74
User Manual 5601.1343.02 ─ 02
6.5.1
Common Commands
The common commands are taken from the IEEE 488.2 (IEC 625-2) standard. A par-
ticular command has the same effect on different devices. The headers of these com-
mands consist of an asterisk "*" followed by three letters.
●
●
●
●
●
●
●
●
●
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the
EVENt
part of
the
QUEStionable
and the
OPERation
registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE
<Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value>
Range:
0 to 255
*ESR?
Event status read
Returns the contents of the event status register in decimal form and subsequently
sets the register to zero.
Return values:
<Contents>
Range:
0 to 255
Usage:
Query only
Remote Control - Commands
www.allice.de
Allice Messtechnik GmbH