Instrument functions
R&S
®
LCX Series
107
User Manual 1179.2260.02 ─ 02
Binning
Indicates the measurement mode that you are working in.
Settings softkey
The (settings) softkey returns to the binning menu, see
<filename>
Indicates the binning
*csv
the measurement is based on.
Counter
Indicates the number of measurements executed.
The R&S
LCX increases the number with each trigger signal received and derives the
statistics from the results and the number of measurements.
You can reset the counter at any time directly on the touchscreen.
Remote command:
Primary - <param>
Indicates the latest measurement value for the first sorting characteristic.
Secondary - <param>
Indicates the latest measurement value for the second sorting characteristic.
Bins
Indicates the configured bins, including the range of each category, and the number of
measured values that fit into one of the categories. After each measurement, the
R&S
LCX highlights the corresponding bin and also assigns a signal at the correspond-
ing pin of the digital output.
EOM ← Bins
Highlighted in red indicates the end of each individual measurement.
OUT ← Bins
Highlighted in red indicates that the measured values of the component do not fit in
any of the configured categories.
-- ← Bins
Unconfigured bin and thus not used for the measurement.
B1 to B6 (B8) ← Bins
Highlights when the measured values of the component fit in one of the configured cat-
egories.
Note:
If you have configured the additional signals "Out of Bin" or "End of Measure-
ment", you can only define the remaining number of categories.
Specific instrument functions