Remote control commands
R&S
®
LCX Series
202
User Manual 1179.2260.02 ─ 02
:LOG:INTerval 0
LOG:COUNt
<Sample Count>[, <Return min or max>]
LOG:COUNt?
[<Return min or max>]
Sets the number of measurement readings in count mode. To set the mode, use the
command
Setting parameters:
<Sample Count>
Range:
1 to 10E+6
*RST:
10
Parameters for setting and query:
<Return min or max> MIN | MINimum | MAX | MAXimum
Example:
See
Example "Logging data settings"
Manual operation:
See
LOG:DURation
<Logging Duration>[, <Return min or max>]
LOG:DURation?
[<Return min or max>]
Defines the duration of logging for the measurement in span and duration mode. To set
the mode, use the command
Setting parameters:
<Logging Duration>
Numeric value in seconds.
Range:
0 to 349199
*RST:
60
Parameters for setting and query:
<Return min or max> MIN | MINimum | MAX | MAXimum
Example:
See
Example "Logging data settings"
Manual operation:
See
LOG:FNAMe
<Logging File Name>
Sets the file name and path for the storing the data recorded during data logging.
LOG subsystem