Common Analysis and Display Functions
R&S
®
FSW
583
User Manual 1173.9411.02 ─ 43
Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
.......................................................583
Predefined Trace Settings - Quick Config
................................................................... 586
Trace 1/ Trace 2/ Trace 3/ Trace 4 (Softkeys)
.............................................................586
Trace 1/Trace 2/Trace 3/Trace 4/Trace 5/Trace 6
Selects the corresponding trace for configuration. The currently selected trace is high-
lighted.
For details see
Chapter 9.5.1.3, "How to Configure a Standard Trace"
Remote command:
Selected via numeric suffix of:
TRACe<1...6>
commands
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
Trace Mode
Defines the update mode for subsequent traces.
For details, see
"Analyzing Several Traces - Trace Mode"
"Clear/ Write"
Overwrite mode (default): the trace is overwritten by each sweep.
The "Detector" is automatically set to "Auto Peak".
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
FSW saves each trace point in the trace memory
only if the new value is greater than the previous one.
The "Detector" is automatically set to "PositivePeak".
This mode is not available for statistics measurements.
Trace Configuration