Measurements and Results
R&S
®
FSW
239
User Manual 1173.9411.02 ─ 43
Figure 7-44: SEM results for multiple sub blocks
Retrieving results via remote control
The measurement results of the spectrum emission mask
test
can be retrieved using
the
CALC:LIM:FAIL?
command from a remote computer; see
on page 1224 for a detailed description.
The
power
result for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? CPOW
;
The
peak
power for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? PPOW
, see
CALCulate<n>:MARKer<m>:FUNCtion:
The measured power
trace
can be queried using
TRAC:DATA?
and
TRAC:DATA:X?
,
The measured
peak power list
can be queried using
TRAC:DATA? LIST
, see
7.6.4
SEM Basics
Some background knowledge on basic terms and principles used in SEM measure-
ments is provided here for a better understanding of the required configuration set-
tings.
................................................................................. 240
Limit Lines in SEM Measurements
........................................................................242
......................................................................................244
Multi-Standard Radio (MSR) SEM Measurements
............................................... 246
SEM with Multiple Sub Blocks ("Multi-SEM")
........................................................ 246
Spectrum Emission Mask (SEM) Measurement