Remote Commands
R&S
®
FSW
1265
User Manual 1173.9411.02 ─ 43
Parameters:
<Frequency>
Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S
FSW
uses the next available frequency. Example: a frequency of
20
MHz is rounded up to the next available frequency (25
MHz).
*RST:
50 MHz
Default unit: Hz
Manual operation:
See
DIAGnostic:SERVice:INPut:PULSed:MCFRequency
<Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
*RST:
7 GHz
Default unit: Hz
Example:
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
DIAGnostic:SERVice:INPut:PULSed:WBFRequency
<Frequency>
Defines the frequency of the internal broadband calibration signal to be used for IF fil-
ter calibration.
This command is only available if the bandwidth extension option R&S
FSW-B160 is
installed.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
2 MHz | 4 MHz | 8 MHz | 16 MHz
If you define a frequency that is not available, the R&S
FSW
uses the next available frequency.
*RST:
16 MHz
Default unit: Hz
Example:
DIAG:SERV:INP:PULS:WBFR 8 MHz
Defines a calibration signal frequency of 8 MHz.
Example:
DIAG:SERV:INP:SEL WBC
DIAG:SERV:INP:PULS:WBFR 4MHz
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
<Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth>
NARRowband | BROadband
Configuring the R&S
FSW