Remote Commands
R&S
®
FSW
1068
User Manual 1173.9411.02 ─ 43
This command is only available for instrument modelsR&S
FSW43/50/67, for fre-
quency sweeps in the Spectrum application.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation:
See
INPut<ip>:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
for the R&S
FSW after an overload condition occurred and the protection mechanism
intervened. The error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
(See
STATus:QUEStionable:POWer[:EVENt]?
Tus:QUEStionable:POWer Register"
The command works only if the overload condition has been eliminated first.
For details on the protection mechanism see
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Example:
INP:ATT:PROT:RES
INPut<ip>:CONNector
<ConnType>
Determines which connector the input for the measurement is taken from.
Suffix:
<ip>
.
1 | 2
For R&S
FSW85 models with two RF input connectors:
1: Input 1 (1
mm [RF Input] connector)
2: Input 2 (1.85
mm [RF2 Input] connector)
For all other models:
irrelevant
Parameters:
<ConnType>
RF
RF input connector