Remote Commands
R&S
®
FSW
1058
User Manual 1173.9411.02 ─ 43
Example:
SWE:EGAT ON
Activate gating
SWE:EGAT:CONT:STAT ON
Activate continuous gating
Manual operation:
See
[SENSe:]SWEep:EGATe:HOLDoff
<DelayTime>
This command defines the delay time between the gate signal and the continuation of
the measurement.
Note:
If you perform gated measurements in combination with the IF Power trigger, the
R&S
FSW ignores the holding time for frequency sweep, FFT sweep, zero span and
I/Q mode measurements.
Parameters:
<DelayTime>
Range:
0 s to 30 s
*RST:
0 s
Default unit: S
Example:
SWE:EGAT:HOLD 100us
Manual operation:
See
[SENSe:]SWEep:EGATe:LENGth
<GateLength>
This command defines the gate length.
Parameters:
<GateLength>
Range:
125 ns to 30 s
*RST:
400μs
Default unit: S
Example:
SWE:EGAT:LENG 10ms
Manual operation:
See
[SENSe:]SWEep:EGATe:LEVel:RFPower
<GateLevel>
Defines the gate level for which the gate is open. Note that any RF attenuation or pre-
amplification is considered when the trigger level is analyzed. If defined, a reference
level offset is also considered.
The input signal must be between 500
MHz and 8
GHz.
This command is only available for triggered gated measurements (
MAN
).
Parameters:
<GateLevel>
For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST:
-20 dBm
Default unit: DBM
Setting Basic Measurement Parameters